What you expect What you Got
| Disease | Medicine |
| Active sites within the column | Test with standard test mixture. If OK add competing base or acid modifier |
| Wrong pH | Correct |
| Wrong column | Change |
| Void volume at inlet | May need repacking |
| Dissolve sample in mobile phase |
Peak fronting
| Disease | Medicine |
| Column overload | See this (Retention Times are not constant) |
| Wrong pH | Correct |
| Sample solvent incompatible with mobile phase | See this (Peak Tailing) |
| Void volume at inlet | May need repacking |
| Wrong sample solvent | Dissolve sample in mobile phas |
Rounded peaks
What you Expect What you Got
| Disease | Medicine |
| Detector outside linear dynamic range | Reduce sample |
| Gain too low | Adjust |
| Column overloaded | See this (retention times are not constant) |
| Time constants (detector, recorder) too high | Reduce |
| Wrong sample solvent | Dissolve sample in mobile phase |