What You Expect What You Got
Disease | Medicine | |
Fluctuation of column temperature | Stabilize. Use column oven | |
Contamination of mobile phase/ | | Use HPLC grade solvent. Degas |
Air in the detector cell | Clean cell. If necessary use a pressure restrictor at outlet | |
Plugged detector outlet line | Replace | |
Default mixing | Check mixer unit. Check flow rate and composition | |
Plugged detector outlet line | Replace | |
Strongly retained materials | Flush column with strong solvent | |
Unoptimized detection | Optimize detector |
Base line noise
What You Expect What You Got
Disease | Medicine |
Air (mobile phase, detector cell, pump) | Flush system, prime pumps, degas mobile phase |
Pump pulse | Use a pulse damper |
Incomplete mixing | Promote complete mixing |
Electronic | Check electronic equipment in the same line |
Leak | Please See No Peaks Or Very Small Peaks |
Disease | Medicine |
Leak | Please See No Peaks Or Very Small Peaks |
Bad quality mobile phase | Use HPLC grade solvents |
Electronics | Locate problem.Get servicing |
Cell contaminant | Clean detector cell |
Weak lamp | Replace detector lamp |
Broad peaks
What You Expect What You Got
Disease | Medicine |
Altered mobile phase | Make new |
Low flow rate | Increase |
Leak | Please See No Peaks Or Very Small Peaks |
Incorrect detector settings | Check and correct |
Column overload | Please see Retention times are not constant |
Void volume. Tubing too long or to wide | Use 0.010" tubing. Shorten path |
Low buffer concentration | Increase |
Column or guard column contamination | Replace |
Void volume at inlet | Repack |
Change in height
Disease | Medicine |
Sample deterioration | Use fresh sample |
Leak | Please See No Peaks Or Very Small Peaks |
Non-reproducible sample volume | Ensure loop is completely filled |
Low detector response | Check detector settings and operating conditions |
Negative peaks
What You Expect What You Got
Disease | Medicine |
Recorder connections | Check polarity |
Refractive index of mobile phase higher than that of solute | Change mobile phase |
Vacancy peaks | Originate from great difference in composition between sample solvent and mobile phase. Dissolve sample im mobile phase |
Mobile phase more absortive than sample components | Use mobile phase that is transparent at the wavelength used |
Ghost peaks
What You Expect What You Got
Disease | Medicine |
Contamination of injector or column | Always flush injector after each injection. |
Retained compound from previous injection | Flush colunm with strong solvent after operation to remove late eluting compounds |