What you expect What you Got
Disease | Medicine |
Active sites within the column | Test with standard test mixture. If OK add competing base or acid modifier |
Wrong pH | Correct |
Wrong column | Change |
Void volume at inlet | May need repacking |
Dissolve sample in mobile phase |
Peak fronting
Disease | Medicine |
Column overload | See this (Retention Times are not constant) |
Wrong pH | Correct |
Sample solvent incompatible with mobile phase | See this (Peak Tailing) |
Void volume at inlet | May need repacking |
Wrong sample solvent | Dissolve sample in mobile phas |
Rounded peaks
What you Expect What you Got
Disease | Medicine |
Detector outside linear dynamic range | Reduce sample |
Gain too low | Adjust |
Column overloaded | See this (retention times are not constant) |
Time constants (detector, recorder) too high | Reduce |
Wrong sample solvent | Dissolve sample in mobile phase |