What You Expect What You got
Disease | Medicine |
Detector off | Check Detector |
Broken connections to recorder | Check connections |
No flow | Check flow |
No sample/Wrong sample | Check sample. Be sure it is not deteriorated. Check for bubbles in the vials |
Wrong settings on recorder or detector | Check attenuation. Check gain |
Disease | Medicine |
Pump off | Start pump |
Flow interrupted | Check reservoirs. Check position of the inlet tubing Check loop for obstruction or air Check degasing of mobile phase Check compatibility of the mobile phase components |
Leak | Check fittings Check pump for leaks and precipitates Check pump seals |
Air trapped in the system | Disconnect column and prime pump Flush system with 100% methanol or isopropanol Contact servicing if necessary |
Reference:
Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa
No comments:
Post a Comment