Disease | Medicine |
Leak | Check for loose fittings Check pump for leaks, salts and noise Check seals |
Mobile phase interrupted | Check reservoirs. Check position of the inlet tubing Check loop for obstruction or air Check degasing of mobile phase Check compatibility of the mobile phase components |
Air trapped in pump | Check flow disconnect column and prime pump Flush system with 100% methanol or iso propanol Contact servicing if necessary |
Worn pump seals | Replace seals Check pistons and replace if necessary |
Higher Pressure than anticipated
What you Expect. What you got
Disease | Medicine |
Pump, injector, tubing | Disconnect column, run pump at 25 mL/min: Is the pressure minimal? Go to next step If pressure still high check systematically from detector to pump for obstruction |
If guard column is obstructed, replace it If the analytical column is obstructed reverse it and flush disconnected from the detector. Use adequate restoration procedures Change inlet frit Replace de column |
reference: