Showing posts with label HPLC Doctor. Show all posts
Showing posts with label HPLC Doctor. Show all posts

Tuesday, December 21, 2010

Base line drift, Base line noise, Broad peaks, Change in height, Negative peaks, Ghost peaks in HPLC

Base line drift



 What You Expect                                                         What You Got



Disease
Medicine
Fluctuation of column temperature
Stabilize. Use column oven
Contamination of mobile phase/

Use HPLC grade solvent. Degas
Air in the detector cell
Clean cell. If necessary use a pressure restrictor at outlet
Plugged detector outlet line
Replace
Default mixing
Check mixer unit. Check flow rate and composition
Plugged detector outlet line
Replace
Strongly retained materials
Flush column with strong solvent
Unoptimized detection
Optimize detector



Base line noise





What You Expect                               What You Got

Disease
Medicine
Air (mobile phase, detector cell, pump)
Flush system, prime pumps, degas mobile phase
Pump pulse
Use a pulse damper
Incomplete mixing
Promote complete mixing
Electronic
Check electronic equipment in the same line
Leak

Please See No Peaks Or Very Small Peaks





 What You Expect                                 What You Got





Disease
Medicine
Leak

Please See No Peaks Or Very Small Peaks

Bad quality mobile phase
Use HPLC grade solvents
Electronics
Locate problem.Get servicing
Cell contaminant
Clean detector cell
Weak lamp
Replace detector lamp





Broad peaks




What You Expect                                 What You Got



Disease
Medicine
Altered mobile phase
Make new
Low flow rate
Increase
Leak
Please See No Peaks Or Very Small Peaks
Incorrect detector settings
Check and correct
Column overload

Please see Retention times are not constant


Void volume. Tubing too long or to wide
Use 0.010" tubing. Shorten path
Low buffer concentration
Increase
Column or guard column contamination
Replace
Void volume at inlet
Repack





Change in height



Disease
Medicine
Sample deterioration
Use fresh sample
Leak
Please See No Peaks Or Very Small Peaks
Non-reproducible sample volume
Ensure loop is completely filled
Low detector response
Check detector settings and operating conditions





Negative peaks




What You Expect                                 What You Got

Disease
Medicine
Recorder connections
Check polarity
Refractive index of mobile phase higher than that of solute
Change mobile phase
Vacancy peaks
Originate from great difference in composition between sample solvent and mobile phase. Dissolve sample im mobile phase
Mobile phase more absortive than sample components
Use mobile phase that is transparent at the wavelength used



Ghost peaks




What You Expect                                 What You Got


Disease
Medicine
Contamination of injector or column
Always flush injector after each injection.
Retained compound from previous injection
Flush colunm with strong solvent after operation to remove late eluting compounds
Reference:

Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa






Monday, December 20, 2010

Peak tailing, Peak Fronting, Rounded Peaks in HPLC chromatogram

Peak tailing



What you expect                                   What you Got



Disease
Medicine
Active sites within the column
Test with standard test mixture. If OK add competing base or acid modifier
Wrong pH
Correct
Wrong column
Change
Void volume at inlet
May need repacking
Dissolve sample in mobile phase



Peak fronting

What you Expect                                What you Got    



Disease
Medicine
Column overload
See this (Retention Times are not constant)
Wrong pH
Correct
Sample solvent incompatible with mobile phase
See this (Peak Tailing)
Void volume at inlet
May need repacking
Wrong sample solvent
Dissolve sample in mobile phas





Rounded peaks




What you Expect                                  What you Got



Disease
Medicine
Detector outside linear dynamic range
Reduce sample
Gain too low
Adjust
Column overloaded
See this (retention times are not constant)
Time constants (detector, recorder) too high
Reduce
Wrong sample solvent
Dissolve sample in mobile phase


Reference:  
Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa

Peak splitting

Peak splitting



What you Expect                                   What you Got





Disease
Medicine
Contamination of column or guard
Remove guard column. If the problem is solved replace it. If not go next
Plugged column
Use adequate recovery procedure. If the problem persists go next
Plugged inlet frit
Replace. If the problem persists discard column

Retention times are not constant

Retention times are not constant



What you Expect                                 What you Got                                         

Disease
Medicine
Leak
Check for loose fittings
Check pump for leaks, salts and noise
Check seals
Change in mobile phase composition
Check. Make new
Air trapped in pump
Check flow
Prime pump. Check and change seals. Be sure the mobile phase is degasse
Temperature fluctuations
Stabilise column temperature. Use column oven
Dilute sample
Sample dissolved in a solvent that is incompatible with the mobile phase
Dissolve sample in the mobile phase whenever possible. Adjust


Reference: 
Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa 

No Pressure or Pressure Lower than anticipated

No Pressure or Pressure Lower than anticipated


                     
What you Expect.                                 What you got





Disease
Medicine
Leak
Check for loose fittings
Check pump for leaks, salts and noise
Check seals
Mobile phase interrupted
Check reservoirs. Check position of the inlet tubing
Check loop for obstruction or air
Check degasing of mobile phase
Check compatibility of the mobile phase components
Air trapped in pump
Check flow
disconnect column and prime pump
Flush system with 100% methanol or iso propanol
Contact servicing if necessary
Worn pump seals
Replace seals
Check pistons and replace if necessary



Higher Pressure than anticipated

  What you Expect.                                 What you got


Disease
Medicine
Pump, injector, tubing
Disconnect column, run pump at 25 mL/min: Is the pressure minimal? Go to next step
If pressure still high check systematically from detector to pump for obstruction
If guard column is obstructed, replace it
If the analytical column is obstructed reverse it and flush disconnected from the detector. Use adequate restoration procedures
Change inlet frit
Replace de column

            
reference: 
Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa 

No Peaks or view Small Peaks in HPLC

No Peaks or view Small Peaks in HPLC
What You Expect                                    What You got


Disease
Medicine
Detector off
Check Detector
Broken connections to recorder
Check connections
No flow
Check flow
No sample/Wrong sample
Check sample. Be sure it is not deteriorated. Check for bubbles in the vials
Wrong settings on recorder or detector
Check attenuation. Check gain



Disease
Medicine
Pump off
Start pump
Flow interrupted
Check reservoirs. Check position of the inlet tubing
Check loop for obstruction or air
Check degasing of mobile phase
Check compatibility of the mobile phase components
Leak
Check fittings
Check pump for leaks and precipitates
Check pump seals
Air trapped in the system
Disconnect column and prime pump
Flush system with 100% methanol or isopropanol
Contact servicing if necessary

Reference:
Prof. Dr. H.J. Chaves das Neves
Universidade Nova de Lisboa 
Related Posts Plugin for WordPress, Blogger...